Product Description
Mini Intelligent Chemical Oxygen Demand Cod Detection Meter
This COD analyzer cod reactor analyzer manual will provide you with the necessary information for the correct use of the instrument. Please read it carefully before using the meter.
Attention :
1.Any liquid or foreign substance entering the inspection chamber may damage the instrument.
2.If the concentration of water samples exceeds the range of the instrument, please dilute the water sample and measure again. The measured value multiplied by the dilution times is the concentration of water sample.
3.There should be no stains, scratches, dust or fingerprints on the surface of the test tube.
Battery requirements: Please use 1.5V AA alkaline batteries.
XC-200 |
Temperature Range | 20-180oC |
Accuracy | ±3 oC(@ 25oC) |
Capacity | 4 vials (the diameter of vial should be not more than 16mm.) |
Warm-up Time | 10-15 minutes, depending on selected temperature |
Heating period | 0-120mins |
Environment temperature | (0-40)oC |
Environment relative humidity | <=85% |
Power Supply | DC 12V/10A |
Dimensions | 110*170*135 mm |
Weight | 1.4 KG with adaptor |
XC-300 |
Temperature Range | 20-180oC |
Accuracy | ±3 oC(@ 25oC) |
Capacity | 4 vials (the diameter of vial should be not more than 16mm.) |
Warm-up Time | 10-15 minutes, depending on selected temperature |
Heating period | 0-120mins |
Environment temperature | (0-40)oC |
Environment relative humidity | <=85% |
Power Supply | DC 12V/10A |
Dimensions | 110*170*135 mm |
Weight | 1.4 KG with adaptor |
Operation processes
COD test
Steps | Operation processes | Description |
1 | Estimate the COD value of the water sample |
2 | Choose the reagent and add the tested water sample according to the corresponding range. |
| COD: o-150mg/L(c11) | COD : 100-1500mg/L(c12) | COD: 1000-15000mg/L(c13) | |
| 1.take 2ml distilled water into a LR reagent tube(ZERO setting) 2.take 2ml water sample into another LR reagent tube. | 1.take 2ml distilled water into a HR reagent tube(ZERO setting) 2.take 2ml water sample into another HR reagent tube. | 1.take 2ml distilled water into a HR reagent tube(ZERO setting) 2. take 1.8ml distilled water into another HR reagent tube ,then add 0.2ml water sample. | ★Note:C12 and C13 can use the same one zero setting tube. ★Danger:There have concentrated sulfuric acid in the test tube,please wear gloves to operate. |
3 | Tighten the reagent tube cap , shake well and put into the reactor, heat for 20 minutes at 165 ºC. | ★Caution: Hot |
4 | After reaction, take out the reagent tube , when cooled to 80 ºC shake reagent tube again. | |
5 | Place the reagent tube on the tube rack and continue to cool to room temperature. This procedure prevents the reagent tube from shaking. | ★Note:The suspended matter needs to be completely precipitated |
6 | Press "(ON/OFF)" button to turn on the meter, press "(SWITCH)" button to select the required test item. | |
7 | Open the detection cap,put it into the zero water sample reagent tube, cover the cap, press the "(ZERO)" button to zero setting. | ★The tube wall should be cleaned. |
8 | Take out the zero water sample reagent tube, put in the water sample reagent tube and cover the cap. | ★Under the c13 , when displayed as XXX mg/L, the detection value is XXX00mg/L. |
9 | Press "(READ)"button.The digits showed in the screen is the value of tested sample. |
10 | Press "(ON/OFF)" button to turn off the meter |
TOTAL PHOSPHORUS TEST
Steps | Operation processes | Description |
1 | Estimate the total phosphorus value of the water sample |
2 | Choose the reagent and add the tested water sample according to the corresponding range. |
Total phosphorus: 0-2mg/L(L22) | Total phosphorus:2-2mg/L(L23) | |
1.take 5ml distilled water into an empty reagent tube(ZERO) 2.take 5ml water sample into another empty reagent tube. | 1.take 5ml distilled water into an empty reagent tube(ZERO) 2.take 0.5ml water sample and 4.5ml distilled water into another empty reagent tube. | ★L22 and L23 can use the same zero reagent tube. |
3 | Add 1 pack reagent 1 to each reagent tube, tighten the cap, shake it up and down for 5 seconds. | ★The reagents doesn't need to be dissolved totally |
4 | Put the reagent tube into the reactor, heat for 15 minutes at 150 ºC. | |
5 | After reaction, take out the reagent tube , when cooled to around 80 ºC shake reagent tube again. | ★Cool 2-3 min |
6 | Press "(ON/OFF)" button to turn on the meter, press "(SWITCH)" button to select the required test item. | |
7 | Open the zero water sample reagent tube, add 1 package reagent 2, shake well, add 7 drops of activator P, shake 5 seconds, put it in the meter, press "(ZERO)" ,the meter will start 1 minute countdown,zero auto setting when countdown finished. Take out the zero water sample reagent tube. | ★The tube wall should be cleaned |
8 | Open the water sample reagent tube, add 1 package reagent 2, shake well until it dissolved totally, add 7 drops activator P, shaking 5 seconds, put it in the meter, press "(READ)",and the meter will be automatically detected after 1 minute countdown. | |
AMMONIA NITROGEN TEST
Steps | Operation processes | Description |
1 | Press "ON/OFF" button to turn on the meter , press "SWITCH" button to select the required test item. | ★Take out the sleeve from the meter ★The marked place face the operator |
2 | Take 10ml distilled water and inject it into the mark line of cell,cover the cap,clean the surface of the cell,then put it into the detection chamber,cover the cap. |
3 | Press "ZERO". It done if shows the "0.00" on the screen. | ★The tube wall should be cleaned |
4 | Pour out the pure water and take a small amount of tested water samples ,wash the cell. Pour out the wet liquid. | |
5 | Take 10ml (scale line) tested water sample to the cell. | |
6 | Add 6 drops of activator A1, shake well, add 6 drops of activator B2 and shake well. Put the cell in the detection chamber, press "()READ" , and the meter will be automatically detected after 3 minutes countdown. | ★environmenttemperature:15ºC-25ºC |
7 | Press" ON/OFF" to turn off the meter.Then clean the cell. | |
Attention:1.The activator is irritating,wash immediately with plenty of water if touched with the skin.
2.This meter can not test the COD/Total phosphorus/Ammonia nitrogen at same time.
Note:
1.15ºC-25ºC is the best temperature for ammonia nitrogen test. It won't influence test result if
Little reagents undissolved.
2.Any solution or foreign matter inside the detection chamber,may damage the meter.
3.If the concentration of water sample higher than the test range,please choose high test range for testing,or tested after diluted by distilled water,the test result should multiply by the dilution ratio.
4.The surface of test tube should avoid waterlogging,scratch,dust, fingerprint and so on.
5.Take action to the sequence and numbers of activator adding.
6.The reagent bag is easy to tear.
7.The sucker of pipette can not be used crossly.
8.It is normal if there have precipitate in the COD reagent tube.
9.The meter should be placed on the table,it can not be tilt or shake violently when testing in other place.
Address:
3/F Jiaofu Building, Middle No. 928 Songjiang Road, Yinzhou Dis., Ningbo, Zhejiang, China
Business Type:
Manufacturer/Factory
Business Range:
Agriculture & Food, Chemicals, Health & Medicine, Industrial Equipment & Components, Instruments & Meters, Light Industry & Daily Use, Manufacturing & Processing Machinery
Management System Certification:
ISO 9001, FSC
Company Introduction:
Welcome to SINOTESTER BIOLOGICAL
Our website: sinotester.en.made-in-china.com
Our Mobile: 86 13732228065
SINOTESTER established in 2008, we are a professional company that manufactures and exports laboratory instruments and equipments. We have our own factory, with strong R&D center, sales team, customer service team and QC team, which grows very fast and obtains good reputation in the market.
Our major products include: Refractometer, Test paper, pH meter, Electrode, TDS meter, Salinity meter, ORP meter, DO meter, Conductivity meter, Turbidity meter, Color comparison tube, test kit etc. Most of the products have CE certificate. All products can be customized according to buyers′ special requirements.
Our uncompromising emphasis on ethical business practices, creativity and continual improvement has earned it a reputation as "Quality Best, Reputation First". A comprehensive quality control policy has been implemented. From raw material purchase to commodities delivery, each process is strictly under supervision. We will make every endeavor to supply high quality products and best services.
Up to now, we have exported our products to more than 56 countries over the world.
"One step ahead" is corporation′s aim and motive to develop strong and better.
We warmly welcome the friends all over the world contact us and establish friendship and business relationship. SINOTESTER will sincerely cooperate with you and joint hands to create a splendid future!